| Origin category | domestic | application area | comprehensive |
|---|
ZJSJ-G high-efficiency filter leak detector
1、 Product Description:
ZJSJ-G is a manual scanning counting leak detector designed according to the requirements of ISO14644-3 B.6 standard, which is used for leak detection of air inlet filters. ZJSJ-G is easy to use, and its purpose is to check and measure the concentration of aerosols, achieving online leak detection of filters. The blue LCD screen can monitor the concentration changes at the downstream end in real time and observe the particle size range (0.3, 0.5, 1.3, 5, 10 μ m).
Users only need to input the leakage point count value through the three keys of "↑", "↓", and "Scan/Stop" (i.e. confirm key) on the handheld display, and during the scanning process, they can obtain the excessive information of the leakage points.
ZJSJ-G consists of four parts, with a high flow particle counter as the main body, combined with a handheld scanning sampling probe, aerosol generator, and diluter.
ZJSJ-G complies with the relevant requirements of ISO14644-3 B.6, EN1822, IEST-RP-CC001.3, C GMPS, EU GMP.
2、 Performance parameters:
Acceptable aerosols/Test Aerosol: standard configuration for DEHS. (PAO, DOP must be explained in advance and quoted separately)
Particle counter counting concentration/Max. Concentration: 1.5 × 104P/L
Leakage value, use your hand to input the three keys "↑", "↓", and "OK" on the display.
Sampling flow rate/Sampling volume: 28.3L/min
Scanning speed/Speed of Scan: Recommended at 3-5 cm/s
Suggested distance between scanning port and filtering surface: 2cm to 4cm
Local dimensions: length * width * height: 240 * 330 * 280mm
Weight: 7Kg
Power supply: AC220V 50Hz/DC17V (AC/DC dual-use)

3、 ZJSJ-G configuration checklist:
|
Serial number |
Component Name |
quantity |
|
1
|
High efficiency filter leak detector body
(including data analysis system)
|
1 unit
|
|
2
|
aerosol generator
|
1 unit
|
|
3
|
Manual scanning device
|
1 piece
|
|
4
|
Dilution device
|
1 unit
|
|
5
|
DEHS aerosol
|
500ml 2 bottles
|
4、 List of vulnerable and consumable parts
|
Serial number |
Component Name |
Year limit |
cost |
remarks |
|
1
|
Printing paper
|
By usage
|
10 yuan/roll
|
Zhejiang Hujing
|
|
2
|
DEHS aerosol
|
By usage
|
300 yuan/bottle
|
Refers to * purchase
|
5 ZJSJ-G Manual Leak Detector Calibration Cost and Details
|
Serial number |
Component Name |
Year limit |
remarks |
|
1
|
Dilution device
|
1 year
|
300 yuan/year
|
|
2
|
aerosol generator
|
1 year
|
300 yuan/year
|
|
3
|
Host debugging
|
1 year
|
600 yuan/year
|
|
|
| ||
6、 Specific parameters of related equipment:
Aaerosol generator
Aerosol Generator
aerosol generator----The ZJSJ-012 aerosol generator is a portable cold gas aerosol generator independently developed and produced by our company. This aerosol generator is suitable for air flow rates of ≤ 50000m3/h and can produce 0.5-2 × 1013P/h polydispersity aerosol. It can meet the requirements of HEPA filter testing for all small clean units to large clean room air supply systems. ZJSJ-012 produces a continuous and stable aerosol particle, which is easy to add to the solution.
Note: Customers need to configure their own compressed air pump or gas source.
|
Performance indicators: |
|
|
model
|
ZJSJ-012 |
|
Aerosol output volume flow rate
|
1-2.5m3/h
|
|
Aerosol generation rate
|
0.5- 2.5×1013P/h
|
|
Atomization pressure
|
Adjustable between 10Kpa and 80Kpa
|
|
Particle size distribution of aerosol
|
0.02-2um
|
|
Aerosol type
|
DEHS cold state polydispersity
|
|
Dimensions H × W × L
|
Height 200 x Width 250 x Length 370 mm
|
|
Weight (kg)
|
3kg
|
|
External air source (compressed air)
|
Clean and stable gas supply below 0.7Mpa
|
BDilution device
Aerosol Dilution
Dilution device----ZJSJ-010 is a portable diluter that can form a dilution ratio of 70-370. It can meet the requirements of HEPA filter testing for all small clean units to large clean room air supply systems. This diluter can form a stable dilution ratio, effectively protecting the leak detector.
|
Performance indicators: |
|
|
model
|
ZJSJ-010 |
|
Dilution ratio
|
70-370 times
|
|
Counting traffic
|
28.3L/min
|
|
Dilution ratio control
|
Flow regulation method
|
|
Application scope |
|
|
Clean Room
|
|
|
Laminar flow workbench
|
HVAC system
|
|
Biological research safety
|
HEPA filter
|
|
Surgery Ward
|