Introduction to Cold and Hot Shock Test Chamber:
The cold and hot shock test chamber is an essential testing equipment in the metal, plastic, rubber, electronics and other material industries, used to test material structures or composite materials. It can withstand continuous environments of extremely high and low temperatures in an instant, and can detect chemical changes or physical damage caused by thermal expansion and contraction of the sample in a short period of time. The test methods that the cold and hot shock test chamber meets are: GB/T2423.1.2, GB/T10592-2008, GJB150.3 high and low temperature impact test. The high and low temperature impact test chamber is divided into three box and two box types according to the testing requirements and standards, with the difference being the testing method and internal structure. The three box type is divided into a cold storage chamber, a heat storage chamber, and a testing chamber. The product is placed in the testing chamber during testing. The two box type is divided into a high temperature chamber and a low temperature chamber, which switch between high and low temperatures by driving the basket with a motor. The product is placed in the basket and moves together with it.
thermal shock chamberFeatures:
1. The controller adopts a 10.2-inch color Chinese English LCDtouch screen,Switching between Chinese and English pages,Easy to operate;
2. The impact method is to switch the air path and import the temperature into the testing area for conducting cold and hot impact tests;
3. The equipment is divided into three parts: high temperature zone, low temperature zone, and testing zone, with insulation structures used between each zone;
5. The cooling system adopts a binary refrigeration system, which has a fast cooling effect. There are two types of cooling methods: air-cooled or water-cooled, and air-cooled is commonly used;
6. Support USB export for convenient data storage;
thermal shock chamberTechnical Specifications:
model |
DJTS-60 |
DJTS-80 |
DJTS-100 |
DJTS-120 |
DJTS-150 |
|
internal dimensions (W×H×D)cm |
50×30×40 |
50×40×40 |
50×40×50 |
60×40×50 |
60×50×50 |
|
external dimensions (W×H×D)cm |
160×175×160 |
160×185×160 |
160×185×170 |
170×185×170 |
170×195×170 |
|
High temperature zone Energy storage temperature range |
RT (room temperature)~+200 ℃ |
||||
|
low-temperature zone Energy storage temperature range |
RT (room temperature)~-78 ℃ |
||||
|
test area Experimental temperature range |
Hot shock experiment (+60 ℃~+150 ℃) |
||||
Cold flushing experiment (A: 0 ℃~-40 ℃/B: 0 ℃~-50 ℃/C: 0 ℃~-60 ℃) | |||||
Impact method |
Test sample stationary (pneumatic cylinder air door switching high/low temperature box) |
||||
Preheating time in high temperature zone |
RT~180 ℃ about 35 minutes |
||||
Pre cooling time in low-temperature zone |
RT~-78 ℃, approximately 105 minutes |
||||
Impact recovery time |
Less than 5 minutes |
||||
Temperature control accuracy |
±0.5℃ |
||||
Temperature Uniformity |
±2.0℃ |
||||
Freezing system |
Original European and American imported fully enclosed or semi enclosed compressor, environmentally friendly refrigerant (R404A/R23) |
||||
texture of material |
inner box SUS#304 Mirror stainless steel plate, outer box SUS#304 Stainless steel plate or baked paint. |
||||
Insulation material |
Imported high-strength fireproof PU foam insulation material |
||||
Accessories |
Lead hole: 1 piece of ¢ 50mm/stainless steel SUS # 304 board storage rack (second layer) |
||||