Main technical parameters:
|
range |
increment |
Measurement volume (μ m)l) |
Maximum allowable system error |
Maximum allowable random error |
||
|
|
|
|
% |
Ml |
% |
Ml |
|
0.5-10Ml |
0.1Ml |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
|
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||
|
1 |
4% |
0.04 |
4% |
0.04 |
||
|
5-50Ml |
0.5Ml |
50 |
1% |
0.5 |
0.50% |
0.25 |
|
25 |
1.50% |
0.375 |
1% |
0.25 |
||
|
5 |
3% |
0.15 |
2% |
0.1 |
||
|
50-300Ml |
5Ml |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
|
150 |
1% |
1.5 |
0.50% |
0.75 |
||
|
50 |
1.50% |
0.75 |
0.80% |
0.4 |
||