Application areas: Boiler Water Power Semiconductor Pharmaceutical Pulp and Paper Chemical/Petrochemical
Functional features:
1. Lower reagent consumption, only 0.2 mL of each reagent is required for each analysis
2. Wider detection range: 0.2ppm-50ppm, automatic range selection
3. Zero point measurement is performed for each analysis cycle to eliminate background interference
For low-temperature water samples, they can be heated to 25 degrees Celsius to ensure complete chemical reactions and improve analysis accuracy
5. The light source used is a high-performance LED lamp with a long lifespan, no need for calibration, and no complex components such as filters or fibers. The system has good stability and high reliability
6. The peristaltic pump used has high reliability, special pump tube material, unique rotor design, and only needs to replace the pump tube every 6 months, making maintenance simple
7. Shutdown self start function. If there are no other button operations after half an hour of pressing the STOP button, the phosphorus meter will automatically start the analysis program (this function can also be cancelled)
8. A large, backlit LCD display screen that can display more measurement, status, and other information
9. Can store calibration and analysis results for 31 days, including detailed information
10. Sampling analysis is available for on-site sampling and testing at any time
11. Compact structural design, convenient for panel or wall installation
12. The enclosure has a high protection level of NEMA4, which prevents dust and dirt, effectively protects the peristaltic pump, and has a long maintenance cycle
13. Wide range, automatic detection of power supply voltage, can accept 100-240 VAV 50-60 Hz
| Product Technical Parameters | ||
| Measurement parameters | measuring range | 0.2 — 50.0 ppm PO43- |
| accuracy | ± 5% or ± 0.5 ppm of the reading, whichever is greater | |
| resolution | 0.1 ppm, full scale | |
| response time | Less than 10 minutes per analysis | |
| repeatability | ± 3% or ± 0.1 ppm of the reading, whichever is greater | |
| detection limit | 0.2ppm | |
| method | Spectral absorption at 470nm wavelength | |
| Operating environment | ambient temperature | 5 — 45℃ (41 — 113℉) |
| ***Humidity | 90% @40℃ | |
| Water sample requirements | Water sample flow rate | 50-1000 mL/min |
| Water sample pressure | ***5 psi | |
| Water sample requirements | continuous | |
| Water temperature range | 5 — 45℃ | |
| Suspended particles | <60 microns | |
| Inlet/outlet connection | 1/4 inch outer diameter ductile pipe - polypropylene pipe or similar material | |
| drain | 3/8-inch outer diameter ductile pipe - polypropylene pipe or similar material | |
| water sample | 1 Road | |
| structure | degrees of protection provided by enclosure | NEMA12 (Instrument Electrical Section) NEMA4X (flow path section) |
| Product Technical Parameters | ||
| structure | External dimensions | 27.5 inches x 16 inches x 6.4 inches (698 mm x 406 mm x 162 mm) |
| Transport weight | 18 kg | |
| electrical | Power requirements | 100 - 240V AC, 110 watts, 50/60Hz |
| Data and Control | Analog Output | 2-channel 4-20 mA. * * * Load 900 ohms |
| relay | 4 SPDTs, 2A @ 250VAC, programmable | |
| digital communication | RS-232 ASCII protocol | |
| authentication | Safety regulations | CE: EN/ IEC61010-1, cTUVus |
| Electromagnetic compatibility (EMC) | CE: EN61326-1 | |
| FCC | Class A | |
| Ordering Information | ||
| order No | describe | |
| Basic must-have configuration | 229500 | 2295 online phosphorus meter |
| 2295REC | Phosphate reagent, 45 day dosage | |
| consumables | 2295REC | Phosphate reagent, 45 day dosage |
| 2295MK | Half year maintenance package | |
| 229510 | 10ppm phosphate standard solution | |