one Overview:
This product is designed and produced according to user needsStepless speed regulation, speed digital display,Equipped with timing and speed regulation. The motor adopts permanent magnet DC motor, which has the characteristics of low noise, obvious speed regulation, and huge low-speed operating force. The stirring rod and sink are both made of stainless steel, enhancing their corrosion resistance. Widely applicable to scientific research, environmental protection, health, petroleum, metallurgical water plants, as well as laboratories in various colleges and universities.
two Product Introduction:
JJ - 6The Type 6 asynchronous automatic lifting electric mixer is driven by a single motor, and the six motors can be used simultaneously. When in use, the height of the motor and mixing rod can be adjusted as needed.
three Instructions for Use:
1When using this instrument, first check if all the accessories accompanying the machine are complete, then install the mixing rod on each of the six shafts, tighten them, and turn off all the work switches. Adjust the speed control knob to the * * * low position, place the solution bottle to be stirred on the machine base, turn on the power, close the power switch, and adjust the lifting platform,Adjust to the appropriate height and align the stirring rod with the center. Turn the timer knob to the normally open position(ON)Turn on the motor power at the desired location or timing,Gradually adjust from low speed to high speed to your desired working speed.
2Before powering on, water should be injected into the pot above the partition. During use, the heating tube must not be exposed to the water surface, otherwise the heating tube will burn or even burst, or the solder will melt and leak, causing electric shock, etc.
3set temperature: PressThe SET key can be used to set or view the temperature setpoint. Press the SET key and the digital display character will start flashing, indicating that the instrument has entered the set state. Press the △ key to increase the set value, press the ▽ key to decrease the set value. Long press the △ key or ▽ key to quickly change the data. Press the SET key again to return the instrument to normal working state and complete the temperature setting. If the temperature controller is intelligent, a separate temperature control manual should be attached
This machine has two indicator lights for heating and constant temperature to indicate the working status, and the temperature gauge shows the actual temperature inside the pot.
4If needed100℃When using boiling water distillation, do not add too much water to prevent overflow during boiling.
four Precautions:
1Using at medium speed within the speed range can extend the lifespan of the instrument.
2Before using the instrument, place the speed control knob in the * * * small position.
3Connect the power and slowly adjust the speed control knob to the desired speed
4Before each shutdown, the timer knob must be turned toPlace the "zero" speed control knob in the * * * small position.
five Technical Indicators:
|
indicator |
JJ - 6 |
speed |
start~2000turn/divide |
power |
2000W |
power supply |
220V/50HZ |