This product is easy to measure, and the sample to be tested does not require any pre-treatment. Simply insert the water sample lifting tube directly into the system water sample to measure the total nitrogen concentration. The device has a particularly high measurement range of 0-500mg/L TN, and this method is mainly used for online automatic monitoring of total nitrogen concentration in waste (sewage) discharge point sources, surface water, etc.
Key Features:
·The unique design enables this product to have lower failure rates, lower maintenance requirements, lower reagent consumption, and higher cost-effectiveness compared to similar products.
·Select valve component: Choose the timing for reagent sampling.
·Measurement component: The reagent measurement is achieved through a visual photoelectric system, which overcomes the quantitative error caused by wear of the peristaltic pump tube; At the same time, quantification of trace reagents was achieved, with each dose being only 1.5 ml, greatly reducing the amount of reagents used.
·Sample injection component: Peristaltic pump negative pressure suction, there is always an air buffer zone between the reagent and the pump tube to avoid corrosion of the pump tube.
·Sealed digestion component: High temperature and high pressure digestion system accelerates the reaction process and overcomes the corrosion of equipment caused by the volatilization of corrosive gases in open systems.
·Reagent tube: Imported modified PTFE transparent hose with a diameter greater than 1.5 mm is used to reduce the probability of particle blockage in water samples.
Technical Indicators:
Method basis: Resorcinol spectrophotometric method;
Measurement range: 0.0-10mg/L, 0.5-100 mg/L, 5-500 mg/L;
Indication error: ≤ 0.5mg/L ± 0.05mg/L; ≥ 0.5mg/L, not exceeding ± 10%;
Repeatability: not exceeding ± 5%;
Stability: not exceeding ± 10% within 24 hours;
Measurement cycle: The measurement cycle is 30 minutes, and the digestion time can be arbitrarily modified between 5 minutes and 120 minutes;
Sampling period: time interval (adjustable from 10 min to 9999 min) and whole point measurement mode;
Calibration cycle: adjustable at any interval and time from 1 to 99 days;
Maintenance cycle: usually once a month, about 30 minutes each time;
Reagent consumption: less than 0.5 yuan/sample;
Output: 4~20mA (2 channels) RS232, RS485;
Environmental requirements: Indoor with adjustable temperature, recommended temperature range of+5-28 ℃; Humidity ≤ 90% (non condensing);
Power supply: AC220 ± 10% V, 50 ± 10% Hz, 5A;
Dimensions: Height 1570 x Width 500 x Depth 450 mm
Other: Abnormal alarms and power outages will not result in data loss; Touch screen display and command input;
After abnormal reset and power failure, the instrument will automatically discharge residual reactants and restore working status.