1、 Product Usage:
The WS-05 water quality analyzer mainly measures COD in sewageCrFive parameters: total phosphorus, ammonia nitrogen, suspended solids, and total nitrogen, CODCrAdopting rapid digestion method, the operation process is simple, fast, and economical, and the measurement results have good comparability with traditional titration method; The total phosphorus is determined using the ammonium molybdate spectrophotometric method specified in the national standard. Potassium persulfate is used to oxidize all phosphorus in the water sample to orthophosphate, which is accurate and meets the requirements of water quality testing; Ammonia nitrogen is measured using Nessler's reagent spectrophotometry. When the tested water sample is added with the reagent, the water sample turns yellow, and the instrument reads the value of ammonia nitrogen based on the depth of the yellow color. This instrument is suitable for water quality testing in large, medium, and small sewage treatment plants, and can be widely used for sewage testing in industrial, chemical, environmental protection, and pharmaceutical industries. It is a commonly used laboratory instrument.
2、 Technical parameters:
Serial number |
display |
measuring range |
Corresponding reagents |
Minimum indication value |
Indication error |
1 |
COD |
150-500.0mg/L |
B,C,D |
0.1mg/L |
±5%Fs |
0-150 mg/L | |||||
500-2500.0mg/L | |||||
2 |
Total phosphorus |
0-5.00mg/L |
Ⅰ,Ⅱ,Ⅲ |
0.01mg/L |
±5%Fs |
3 |
Ammonia nitrogen |
0-50.00mg/L |
Ⅰ,Ⅱ |
0.01mg/L |
±5%Fs |
4 |
Suspended solids |
0-200.0mg/L |
0.1mg/L |
±5%Fs |
|
0-1000.0mg/L | |||||
5 |
Total nitrogen |
0-100.0mg/L |
Ⅰ,Ⅱ,Ⅲ |
0.1mg/L |
±5%Fs |
Repeatability: ≤ 2%
Power Supply: AC 220V 50Hz
Instrument calibration
The instrument has been calibrated at the factory and can be used directly. If the user needs to recalibrate, please provide a standard solution by yourself.
3、 Product features:
1. Time saving, each parameter detection takes less than 1 hour.
2. It can automatically zero and calibrate from 1 to 5 points.
3. Standard optical glass sample slots have stronger interchangeability.
4.999 measurement storage function.
5. Date display function, storing one date and time at a time for easy querying.
6. Experimental results can be printed.
7. Full Chinese menu, easy to operate and understand
8. The instrument has a power on self-test function. Please be patient and wait for the self-test to complete before operating.