Instrument Introduction:
One Overview
This product is a research achievement of Zhejiang University and is at the forefront of domestic testing technology indicators and performance. Won the invention patent issued by the China National Intellectual Property Administration (patent number:ZL 97 1 00943.0). It is a domestically produced scientific instrument recommended by the China Instrumentation Association and the Analysis and Testing Association.
A .Main purpose and scope of application
Capacity titration Karl Fischer(Karl Fischer)Method is an accurate method for determining the moisture content of substances in various chemical methods, and has become a commonly used analytical method internationally.ZKF1Karl Fischer water analyzer is a sophisticated automatic analysis system that uses Karl Fischer volumetric titration to determine water content. It is a very suitable conventional analytical instrument.
This instrument can be used in principle for the determination of moisture content in drugs, organic chemicals, inorganic chemicals, petrochemicals, fertilizers, pesticides, dyes, food, beverages, surfactants, cosmetics, fibers and other objects.
Due to the diverse characteristics of the tested substance, a solvent that can dissolve the substance should be selected for actual measurement.
B . Product Features
⑴ High measurement accuracy
·Determine and display the reaction endpoint based on the polarization voltage of the dual platinum electrodes, and titrate at the endpoint.
·Using finely manufactured measuring tubes, with a capacity error of ≤±0.004
·Fully enclosed system to prevent the infiltration of environmental moisture.
⑵ Wide measurement range
0.001%–100 %Moisture content.
II technical performance
A.Main performance
•Principle of Karl Fischer volumetric method for moisture determination
•Measurement of polarization voltage using dual platinum electrodes and titration endpoint
•Three and a half peopleLEDDigital display unitml
•Standard arsenic or arsenic free reagents can be used
•Electromagnetic mixer with stepless speed regulation
•Special bottle connector for convenient connection with standard reagent bottles and solvent bottles
•Electric diaphragm pump controls solvent injection and waste liquid discharge
B.Main parameters
•measurement range 0.001 %–100 %water content
•resolution 0.01ml
•Nominal total capacity of burette 20 ml
•Repeatability of water titration: ≤0.01
•Linear correlation coefficient of water titration:≥0.998
III Size and weight
A.Overall dimensions:
•host 34 × 34 × 30 (cm)
•diaphragm pump 21 × 12 × 6.5 (cm)
B.Weight:
host 9.5 kg